TiePie Handyscope HS6 DIFF

TiePie Handyscope HS6 DIFF
PartNumberHS6
Delivery time4-5 weeks
Price from937,00 ex. VAT

The new 1 GS/s 14 bit quad differential channel High Resolution Oscilloscope.

This powerful USB 3.0 super speed oscilloscope combines fast sampling up to 1 GS/s with high resolutions of 12, 14 and 16 bit and a large memory of 64 MSamples on all four channels. The oscilloscope supports continuous streaming measurements up to 200 MS/s and can be synchronized with other oscilloscopes using the CMI interface to form a multi channel combined instrument with synchronized time base. The CMI interface is available by default on the Handyscope HS6 DIFF. Optionally, the Handyscope HS6 DIFF can be delivered with SureConnect connection test and resistance measurement on each channel. Also, the Handyscope HS6 DIFF can be delivered with SafeGround option. With SafeGround you can switch the differential inputs of the Handyscope HS6 DIFF into single ended inputs with ground protection. It allows to make measurements using standard attenuating probes and protects the scope when a short circuit to ground is created.

Integration

Bohemian Technology Group s.r.o. is an official technical and technical support partner for integrating TiePie oscilloscopes into all applications. We specialize in the .NET C # implementation and can help you with less common tasks.

 

TiePie Handyscope HS6 DIFF
Number of input channels 4 analog
CH1, CH2, CH3, CH4 Isolated male BNC
Type Differential input
Resolution 8, 12, 14, 16 bit user selectable
DC Accuracy 0.25 % (0.1 % typical) of full scale ± 1 LSB at 20°C to 25°C
To achieve rated accuracy, allow the instrument to settle for 20 minutes. When subjected to extreme temperatures, allow additional time for internal temperatures to stabilize.
Ranges ±200 mV to ±80 V full scale
Coupling AC/DC
Impedance 2 MOhm / 12 pF ± 1 %
1 MOhm / 20 pF ± 1 % when SafeGround enabled
Maximum voltage 200 V (DC + AC peak < 10 kHz)
Maximum Common Mode voltage 200 mV to 800 mV ranges : 2 V
2 V to 8 V ranges : 20 V
20 V to 80 V ranges : 200 V
Common Mode Rejection Ratio -47 dB
Bandwidth HS6D-1000 HS6D-500 HS6D-200 HS6D-100 HS6D-50
-3dB at 75 % of full scale input 250 MHz 250 MHz 250 MHz 100 MHz 100 MHz
Limit, selectable per channel Off (250 MHz)
150 MHz
100 MHz
50 MHz
Off (250 MHz)
150 MHz
100 MHz
50 MHz
Off (250 MHz)
150 MHz
100 MHz
50 MHz
Off (100 MHz)
75 MHz
50 MHz
25 MHz
Off (100 MHz)
75 MHz
50 MHz
25 MHz
AC coupling cut off frequency (-3dB) ±1.5 Hz
SureConnect Optionally available (option S)
Maximum voltage on connection 200 V (DC + AC peak < 10 kHz)
Resistance measurement Optionally available (option S)
Ranges 100 Ohm to 2 MOhm full scale in 1-2-5 sequence
Accuracy 1 % of full scale
Response time (to 95%) < 10 µs
SafeGround Optionally available (option G)
Maximum voltage on connection 200 V (DC + AC peak < 10 kHz)
Maximum switch off current 500 mA
Response time < 100 ns
   
Maximum sampling rate HS6D-1000 HS6D-500 HS6D-200 HS6D-100 HS6D-50
8 bit  
measuring one channel 1 GS/s 500 MS/s 200 MS/s 100 MS/s 50 MS/s
measuring two channels 500 MS/s 200 MS/s 100 MS/s 50 MS/s 20 MS/s
measuring three or four channels 200 MS/s 100 MS/s 50 MS/s 20 MS/s 10 MS/s
12 bit  
measuring one channel 500 MS/s 200 MS/s 100 MS/s 50 MS/s 20 MS/s
measuring two channels 200 MS/s 100 MS/s 50 MS/s 20 MS/s 10 MS/s
measuring three or four channels 100 MS/s 50 MS/s 20 MS/s 10 MS/s 5 MS/s
14 bit 100 MS/s 50 MS/s 20 MS/s 10 MS/s 5 MS/s
16 bit 6.25 MS/s 3.125 MS/s 1.25 MS/s 625 kS/s 312.5 kS/s
   
Maximum streaming rate 1 HS6D-1000 HS6D-500 HS6D-200 HS6D-100 HS6D-50
8 bit  
measuring one channel 200 MS/s 2 100 MS/s 2 40 MS/s 20 MS/s 10 MS/s
measuring two channels 200 MS/s 3 50 MS/s 3 20 MS/s 10 MS/s 5 MS/s
measuring three or four channels 100 MS/s 4 25 MS/s 4 10 MS/s 5 MS/s 2.5 MS/s
12 bit  
measuring one channel 200 MS/s 3 50 MS/s 3 20 MS/s 10 MS/s 5 MS/s
measuring two channels 100 MS/s 4 25 MS/s 4 10 MS/s 5 MS/s 2.5 MS/s
measuring three or four channels 50 MS/s 5 12.5 MS/s 5 5 MS/s 2.5 MS/s 1.25 MS/s
14 bit  
measuring one channel 100 MS/s 3 50 MS/s 3 20 MS/s 10 MS/s 5 MS/s
measuring two channels 50 MS/s 4 25 MS/s 4 10 MS/s 5 MS/s 2.5 MS/s
measuring three or four channels 25 MS/s 5 12.5 MS/s 5 5 MS/s 2.5 MS/s 1.25 MS/s
16 bit 6.25 MS/s 6 3.125 MS/s 1.25 MS/s 625 kS/s 312.5 kS/s
   
Sampling clock source  
Internal TCXO
Accuracy ± 0.0001 %
Stability ± 1 ppm over 0°C to 55°C
Time base aging ±1 ppm/year
External LVDS, on CMI connectors
Input range 10 MHz ± 1 %
16.369 MHz ± 1 %
   
Memory Standard model XM option
8 bit resolution  
Measuring one channel 1 MS per channel 256 MS per channel
Measuring two channels 512 KS per channel 128 MS per channel
Measuring three or four channels 256 KS per channel 64 MS per channel
12, 14, 16 bit resolution  
Measuring one channel 512 KS per channel 128 MS per channel
Measuring two channels 256 KS per channel 64 MS per channel
Measuring three or four channels 128 KS per channel 32 MS per channel
   
Trigger  
System Digital, 2 levels
Source CH1, CH2, CH3, CH4, OR, digital external
Trigger modes Rising edge, falling edge, any edge, inside window, outside window, enter window, exit window, pulse width
Level adjustment 0 to 100 % of full scale
Hysteresis adjustment 0 to 100 % of full scale
Resolution 0.006 % (14, 16 bits) / 0.025% (12 bits)
Pre trigger 0 to selected record length, 1 sample resolution
Post trigger 0 to selected record length, 1 sample resolution
Trigger hold-off 0 to 63 MSamples, 1 sample resolution
Trigger delay 0 to 16 GSamples, 1 sample resolution
Digital external trigger  
Input Extension connector pins 1 and 2
Range 0 to 2.5 V (TTL)
Coupling DC
Jitter ≤ 1 sample

Download:

File
Download datasheet

Product images

 

Related